Altitude Correction
The CDM-6000 is factory calibrated for operation at sea level. When operated
at higher elevations, the calibration may be adjusted to correct for this factor.
The CDDS-2 can be set to the exact operating elevation to eliminate the
measurement error caused by decreased barometric pressure at higher
altitudes. This altitude compensation is done by the CDDS-2 controller. Let
the CDM-6000 stabilize to the ambient CO2 concentration. To change the
operating elevation:
1. Press and hold both the 'SET' and 'SENSOR' buttons for about five (5)
seconds until all the yellow LED's blink. The current elevation setting is
displayed.
2. Use the 'UP' and 'DOWN' buttons to adjust the display reading to the
current elevation in feet. The display moves in 10 foot increments.
3. After several seconds of no button activity the CDDS-2 reverts to its normal
operating mode.
Specifications
Operating principle
Gas sampling method
Measurement range
Typical drift (per year)
Accuracy
Repeatability
Calibration Interval
Response time
Operating temperature
Operating humidity
Storage temperature
Power requirements
Power consumption
Voltage output (linear)
Current output (linear)
Warm-up time
Dimensions
Weight
Digital Display
Operating life expectancy
Warranty
Non-dispersive infrared (NDIR)
Diffusion or sample draw
0 - 5000 ppm CO 2
±75 ppm (@ 1200ppm)
±5% of reading or ±75 ppm
±20 ppm
Five Years
Less than 1 minute
0 to 50 ° C
0 - 90% RH (non condensing)
-30 to + 60 ° C
20 - 28 VRMS AC, 18 - 30 VDC
Less than 2W @ 24 VAC
0 - 10 volts DC standard
4 - 20 mA
(R
< 500 ohm)
L
3 minutes
5.2" x 3.2" x 1.4"
6.5 Oz. (.35 Kg)
4 digit, .35" LCD
10 years typical
90 day, parts and labor
P.O. Box 1318 Gresham Or. 97030 USA
1-800-669-2113 Fax 1-503-663-1147
[email protected] www.greenair.com