Gamma Compact Gebrauchsanweisung - Seite 5

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Produkt für die Atemwege Gamma Compact herunter. Gamma Compact 19 Seiten. Ultrasonic nebulizer system

3. carryInG out nebulIzInG therapy
1. Press ON switch, the green light will turn on and the unit will start to operate. Adjust the air
regulator to the desired intensity.
. The unit will automatically turn off at minutes. If you wish to continue nebulizing, check if the
medication amount in the pool is the suitable one. Add more if necessary. Then press the ON/OFF
switch twice and the unit will start to operate again.
. Once use is over, unplug the equipment and disconnect the AC adapter. Remove the clear hous-
ing and dry the left medication using a clean cloth. There will always remain some solution in the
basin which may not be used (between 0. and 1 ml).

4 cleanInG and dIsInfectIon

warnInG: it is highly recommended to clean the pieces before first use of the device and
after each use
1. Wash all the pieces, except for the air filter, with warm water and detergent. Rinse with warm tap
water to remove the detergent.
. Dip the pieces into a clean bowl containing parts of warm water and 1 part of vinegar during
0 minutes, disinfectant can also be used.
. Wash your hands and take out the pieces from the disinfectant solution, rinse them with warm
tap water and leave them on an absorbent piece of paper.
. Keep the pieces into a hermetic bag. Do not dry them with a towel.
. Clean the outer surface with a dry piece of cloth.
Do not use any volatile fluid like alcohol or benzene.

5. possIble problems and how to solve them

If something wrong happens while the device is working, please check and correct the following
points:
8
www.gamma-medica.com.ua
symptom
possible cause
Although the device is
1. Incorrectly plugged
on, it does not work
. No power
. The AC Adapter is not
connected
The mist does not come
There is no solution into
out
the cup
Little mist is
1. The air filter is dirty
delivered
. There is no enough
solution into the cup
. Little power
. Air excess
The device operates
Timer off
for a period and then
turns off
Should the device not start working properly again, please contact the Service Centre; the address
is stated in the Guarantee Card.

6. technIcal specIfIcatIons

Method of nebulization
Operating Voltage
Class
suggested solutions
1. Check the plug is
properly connected
. Check the power is
not gone out
. Check the AC Adapter
is properly connected
Pour some drug solution
into the medication cup
1. Change the filter
. Pour some drug
solution into the
medication cup
. Increase the power
. Regulate the air flow
Press the key twice and
turn it on again
Ultrasonic Oscillation
0V, 0Hz
www.gamma-medica.com.ua