Gamry Reference 3000 Schnellstart-Handbuch - Seite 2
Blättern Sie online oder laden Sie pdf Schnellstart-Handbuch für Messgeräte Gamry Reference 3000 herunter. Gamry Reference 3000 4 Seiten. Usb potentiostat installation
Auch für Gamry Reference 3000: Schnellstart-Handbuch (16 seiten)
When possible, connect the Chassis Ground on the back of your
potentiostat to a known, good earth ground.
Connect the cell cable to the Calibration side of the UDC4 dummy cell
included with your instrument.
Place the dummy cell inside the Calibration
Shield, close the lid, and connect the black
floating-ground lead of your cell cable to the
Shield's grounding post.
Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument
Choose your potentiostat. Select the radio button
for calibration type Both, and click the OK button.
After you confirm several instructional
messages, the calibration proceeds
automatically, and you are notified if the
calibration is successful.