2.2 Deuterium Light

By switching the Deuterium switch downwards to ON the Deuterium lamp is preheated. After a
warm-up time of 10-20 sec. the Deuterium lamp will ignite automatically.
After successful ignition the Deuterium LED lights up green and indicates that the Deuterium lamp is
in operation. Should the Deuterium lamp not ignite the Deuterium LED lights up red. This indicates a
malfunction of the lamp. See section "4.3 Trouble Shooting". Switch to OFF to reset the Error LED.
Switching the Deuterium switch to upwards to OFF the Deuterium lamp switches off.

2.3 Halogen Light

Switching the Halogen switch downwards to ON, the Halogen lamp switches on. The Halogen LED
indicates the operation mode.
Switching the Halogen switch upwards to OFF, the Halogen lamp switches off.
2.4 TTL Shutter – Open
This switch is used to open and close the shutter either manually or by TTL pulse.
The LED indicated if the shutter open (LED on) or closed.

2.5 Power connector with double Fuse and switch

The power connector for 100-240 VAC is located on the rear of the AvaLight-D(H)-S. Be careful to
use it with a designated power range only, please use the included power cord with the instrument.
For UK, US and Australian power cords, contact Avantes Technical Support.
The 2 Fuses are 1A slow blowing Fuses.
Disconnect power before opening housing or replace Fuse.
The installation category for this equipment is Class 2.
Revision: 1-00
© Avantes BV
Switch position
Deuterium light
Up
OFF
Down
ON (after 10-20 seconds)
Switch position
Up
Down
Position switch
Shutter
Up
Closed or TTL activated
Down
Open
www.avantes.com
Halogen light
OFF
ON