Alicat Scientific EPC Series Betriebshandbuch - Seite 8
Blättern Sie online oder laden Sie pdf Betriebshandbuch für Controller Alicat Scientific EPC Series herunter. Alicat Scientific EPC Series 17 Seiten. Pressure vacuum controller
Analog Signals
Analog Output
Most devices include a primary analog output signal, which is linear over its
entire range. For ranges that start at 0 Vdc, a zero-pressure condition is
indicated at approximately 0.010 Vdc. Full scale pressure is indicated by the
top of the range: 5 Vdc for 0-5 Vdc, 10 Vdc for 0-10 Vdc, and so on.
Serial Communication
Establishing Serial Communication
After connecting your flow controller using a communications cable, you will
need to establish serial communications through a real or virtual COM port
on your computer or programmable logic computer (PLC).
• If you have connected your device to a serial port, note its COM port
number. This can be found in Windows® Device Manager.
• If you have a serial to USB adapter, plug it into your computer, ensure the
proper driver is installed, and note the COM port number.
The controller will be configured with the following settings:
• Baud: 19200 (default; others can be used if the computer, its software
and the controller are all set to the same rate)
• Data bits: 8
• Parity: none
• Stop bits: 1
• Flow control: none
Alicat devices use ASCII commands which are easily used in a serial terminal
or integrated into a PLC.
✓
Note: In what follows,
hexadecimal D). Serial commands are not case-sensitive.
indicates an ASCII carriage return (decimal 13,
8