4.1 Removal Process
1.
Remove the battery cover assembly (17), and then remove the battery (11).
2.
Gently place the 4024 series spectrum analyzer on the table with back upwards, and remove eight
fixing screws (four on the back and four on the side) of the casing with a Phillips screwdriver.
3.
Separate 4024 front casing from its rear casing (Attention: In this case, two parts of the casing are
connected by multiple cables, do not separate the casing forcibly). Firstly, remove one end of the
interconnecting cable between the front and rear casings, and then place them on the table. In case
of reassembly, pay attention to the relationship of the connection cables, as shown in the following
figure.
Fig. 2 Removal of Battery
Fig. 3 Removal of 6 Fixing Screws