AB Sciex API 5000 LC/MS/MS Hardware-Handbuch - Seite 15

Blättern Sie online oder laden Sie pdf Hardware-Handbuch für Laborausrüstung AB Sciex API 5000 LC/MS/MS herunter. AB Sciex API 5000 LC/MS/MS 16 Seiten.

AB Sciex API 5000 LC/MS/MS Hardware-Handbuch
API 5000
The following table contains generic parameters for the API 5000™ instrument.
The first number under each scan type is the preset value; the range of numbers
is the accessible range for each parameter.
Table 2-1 API 5000™ Instrument Parameters
Parameter
ID
CUR
CAD
IS
(1)(2)
NC
TEM
OR
(DP = OR)
Q0
(EP = –Q0)
IQ1
(IQ1 = Q0 + offset)
ST
(ST = Q0 + offset)
RO1
(IE1 = Q0 – RO1)
RO1
(IE1 = Q0 + offset)
(1)IonSpray™ ion source (2)TurboIonSpray
and 0=OFF (5)DuoSpray™ ion source, 1=TIS, and 2=HN
Document Number: D1000092206 B
Instrument Parameters
Positive ion mode
Access
ID
Q1
CUR
10
10 to 50
CAD
0
n/a
5500
IS
(1)(2)
0 to 5500 0 to 5500 0 to
3
NC
(3)
(3)
0 to 5
TEM
0
(2)(3)
(2)(3)
0 to 750
DP
120
0 to 400
EP
10
15 to 2
IQ1
Q0 + (–1)
–0.5 to –2 –0.5 to –2 –0.5 to
ST
Q0 + (–7) Q0 + (–7) Q0 +
–8 to –4
IE1
1
0.5 to 2
RO1
n/a
Negative ion mode
Q3
MS/MS Q1
10
10
10
10 to 50
10 to 50 10 to 50 10 to 50 10 to 50
1
4
0
0 to 12
0 to 10
n/a
5500
5500
–4500
–4500 to
5500
0
3
3
–3
0 to 5
0 to 5
–5 to 0
0
0
0
0 to 750
0 to 750 0 to 750 0 to 750 0 to 750
120
120
–100
0 to 400
0 to 400 –400 to
0
10
10
–10
15 to 2
15 to 2
–15 to
–2
Q0 + (–1) Q0 +
Q0 + 1
(–1)
0.5 to 2
–2
Q0 + 7
(–7)
–8 to –4
–8 to –4
4 to 8
n/a
1
–1
0.5 to 2
–2 to
–0.5
Q0 + (–2) n/a
n/a
–0.5 to –2
ion source (3)Heated Nebulizer (4) 1=ON
®
2
Q3
MS/MS
10
10
1
4
0 to 12
0 to 12
–4500
–4500
–4500 to
–4500 to
0
0
–3
–3
–5 to 0
–5 to 0
0
0
–100
–100
–400 to
–400 to 0
0
–10
–10
–15 to
–15 to –2
–2
Q0 + 1
Q0+1
0.5 to 2
0.5 to 2
Q0 + 7
Q0 + 7
4 to 8
4 to 8
n/a
–1
–2 to
–0.5
Q0 + 2
n/a
0.5 to 2
15