Aqua Free Germlyser D Gebrauchsanweisung - Seite 2

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Water Filtration Systems Aqua Free Germlyser D herunter. Aqua Free Germlyser D 2 Seiten.

1.3 Scope of delivery
• Germlyser
®
D
• Transport box
• Cable tie for product return and box fastener
• Zip lock bag for used filter
• Information material, sticker for filter replacement
1.4 Reprocessing
The Germlyser
®
D is a high-quality, reusable
customer at a fixed rental rate. Reprocessing is carried out exclusively by Aqua free.
1.4 Warning notes
Check the packaging for damage prior to use. After unpacking / Prior to first-time use of the
D, the filter must be examined to ensure that it is undamaged.
®
Germlyser
The Germlyser
®
D may only be handled by instructed/trained persons.
Aqua free GmbH does not accept any warranty for use of the Germlyser
life of up to 2 months. The operator is responsible for its use.
®
The Germlyser
D housing must not be opened or altered.
2 Certificates / Declaration of conformity
The company Aqua free GmbH is certified according to DIN EN ISO 13485.
filters are
-labelled medical devices.
The Germlyser
®
Extract from the declaration of conformity
Aqua free GmbH, Winsbergring 31, DE-22525 Hamburg:
®
For the Class I medical product Germlyser
D the manufacturer declares conformity with the basic
requirements of Annex I of council directive 93/42 EEC and its amendment 2007/47/EC as required
by Annex IX of directive 93/42 EEC on medical devices.
3 Labelling
Manufacturer label
D/GB
Product information material
Do not use if package
is damaged
4 Preparation for commissioning and use of product
4.1 Installation of quick-release coupling
- Remove the shower head
- Install the quick-release coupling (including water stop)
Check the adapter for leakage by opening the water tap
without installing the filter!
-labelled medical device that is made available to the
D after the specified service
®
SN
Individual product label
Medical device label
Non-sterile
NON
STERILE
4.2 Filter installation
Please disinfect your hands and put on disposable gloves
before filter installation
- Open the packaging on the adapter side of the filter
- Wet the o-ring with a commercial disinfectant
(for easier and safer installation)
- Leave the filter in its packaging and insert it into the
quick-release coupling with a slight turning movement until
the filter audibly engages
- Align filter and remove packaging
- Open the water supply to test the flow and check for leakage
4.3 Filter replacement documentation
- Note the date of the next filter replacement on the sticker
and place it on the filter in an easily visible position –
but not on the barcode.
4.4 Handling recommendations when used for wound rinsing
- Rinse the filter after use.
- Dry the spray head and housing and disinfect by wiping with
standard alcohol disinfectant wipes.
- Store the filter in a hygienic environment until it is next used for rinsing wounds.
- Rinse the filter before re-using for rinsing wounds.
Note: To further reduce the potential for infection it is recommended that each
filter is dedicated to one individual patient.
5 Removal of used filter / filter change
Please disinfect your hands and put on disposable gloves
before removing filter
- Place the zip lock bag over the filter, loosen the adapter
and pull off the filter
- Place the filter in the provided transport box
- The transport box must be equipped with protective foil
for safe transport
- Close transport box with cable ties for dispatch
- Install a new filter (see point 4)
6 Accessories for Germlyser
®
D
- Quick-release coupling (including seal) with G½" female thread
Item No.: 100573
- Quick-release coupling (including seal) with G½" male thread
Item No.: 100300
- Bacillol Tissues, Disinfecting wipes in dispenser
Item No.: 100423