Aqua Lung CORE Service-Handbuch - Seite 12

Blättern Sie online oder laden Sie pdf Service-Handbuch für Controller Aqua Lung CORE herunter. Aqua Lung CORE 18 Seiten.

Aqua Lung CORE Service-Handbuch
SYMPTOM
nd
Leak or free flow at 2
stage
Insufficient purge flow or work of
breathing too high
Water leak
nd
Service Manual Core 2
Rev. 05/2015
Table 1. Troubleshooting Guide
POSSIBLE CAUSE
1. The MP is too high
2. The rubber seating (129638) is worn or
damaged
3. The valve seat (AP2033) is not correctly
adjusted
4. The lever (129178) is bent
5. The sealing face of the valve seat
(AP2033) is damaged
6. The spring (AP2021) is damaged
1. The MP is too low
2. The valve seat (AP2033) is not correctly
adjusted, the lever adjustment is too low
3. The MP hose is obstructed
4. The lever (129178) is bent
1. There is an hole in the mouthpiece
(123697 or 123698)
2. The diaphragm (129150) is damaged
3. The exhaust valve (129174) is damaged
4. The Venturi lever O-ring (AP1438) is
dirty, worn or damaged
5. The diaphragm is not correctly fitted
between the case (129931 or 129932) and
the washer (129133)
6. The case (129931 or 129932) is
damaged
7. The valve spindle O-ring (124706) is
damaged
stage

APPENDIX

Service Manual Core 2
TREATMENT
1. Refer to First stage Troubleshooting
Guide
2. Replace the rubber seating
3. Readjust the valve seat
4. Replace the lever
5. Replace the valve seat
6. Replace the spring
1. Refer to First stage Troubleshooting
Guide
2. Readjust the lever and the valve
seat
3. Clean or replace the hose
4. Replace the lever
1. Replace the mouthpiece
6. Replace the diaphragm
7. Replace the exhaust valve
4. Replace the O-ring
5. Disassemble the front cover and refit
the assembly correctly
6. Check the sealing face of the
exhaust valve. Replace the case.
7. Replace the O-ring
nd
stage
12