Allied Healthcare EPV200 Benutzerhandbuch - Seite 10
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Medizinische Ausrüstung Allied Healthcare EPV200 herunter. Allied Healthcare EPV200 20 Seiten. Portable battery powered ventilator with oxygen source required
9
Battery Replacement
Loosen
screws on
battery
door in
4 places
The batteries are replaced by loosening the 4 thumbscrews on the battery door located
on the front of the unit. The batteries should be removed by gently prying the top (+) of
the battery up and toward the bottom of the unit. Replace the batteries with 2 D size
alkaline batteries only. Do not mix new and old batteries when installing in unit. The
batteries must be inserted with the positive (+) end of the battery toward the top of the
unit. For ease of installation, install the bottom (-) of the battery first. Replace the
battery door and tighten the 4 thumbscrews.
Turn the unit on and verify that the displays turn on. Connect a 50 psi oxygen source to
the unit and verify that the unit does deliver breaths. The breath delivery alarm will
sound if the unit is not connected to a test lung. This is normal and not a concern. The
ventilator is now ready for use. Always turn the ventilator off when not in use to
preserve battery life. The ventilator does not turn off automatically.
Warning: Do not recharge alkaline batteries.
10
Cleaning
The EPV200 should be cleaned after each use. To clean the EPV200, keep the gas
supply hose on the unit to prevent contamination of the oxygen circuit.
Warning: Cleaning procedures should be performed in an environment
free of oil and petroleum based products.
The EPV200 has been designed to be water resistant, but the unit cannot be
submerged or sprayed down for cleaning.
Wipe the unit with a damp rag containing a mild detergent to remove any residue from
the surface. Once the residue has been removed, the unit may be wiped with isopropyl
alcohol or a cold disinfecting solution to kill bacteria. The unit should then be wiped with
water to remove any film left by the cold disinfecting solution. Make sure the unit is dry
before putting the unit away.
Positive (+) end of battery
towards the top of the unit.
9