Allied Healthcare Schuco S130 Schnellstart-Handbuch - Seite 7
Blättern Sie online oder laden Sie pdf Schnellstart-Handbuch für Medizinische Ausrüstung Allied Healthcare Schuco S130 herunter. Allied Healthcare Schuco S130 12 Seiten.
READ AND SAVE THESE
INSTRUCTIONS
TUBING REPLACEMENT:
个
| THE TUBING SHOULD BE CHANGED AFTER EACH PATIENT USE OR ON A MONTHLY BASIS IF THE SAME
|
/M
| PATIENT USES UNIT.
BACTERIA FILTER REPLACEMENT:
The high efficiency bacteria filter (01-90-3100) is custom engineered to prevent fluid and aerosol contamination fluid and aerosol contamination
of mobile suction units. This filter features a hydrophobic, micro-porous membrane, which filters air with maximum efficiency (0.3-micron
particles in air), while blocking the flow of aqueous fluids and aerosol contaminants. This high efficiency filter protects against suction pump
contamination. in the case of canister/bottle overflow from reaching the pump.
人
THE HYDROPHOBIC BACTERIA FILTER SHOULD BE CHANGED AFTER EACH PATIENT USE ORONA
n
MONTHLY BASIS IF THE SAME PATIENT USES UNIT.
ἲ
CARE MUST BE TAKEN WHEN REPLACING THE FILTER TO PREVENT THREAD DAMAGE DUE TO CROSS.
|
THREADING AND OVER TIGHTENING.
To replace the bacteria filter:
1..
Unscrew the filter from the filter connection fitting by hand turning the filter in a counterclockwise direction.
2.
Carefully screw the new filter into the fitting by turn the filter in a clockwise direction.
3.
Tighten only finger tight.
RECOMMENDED SERVICE PROCEDURE FOR SCHUCO-VAC PRODUCTS:
1.
Preventative Maintenance:
Preventative maintenance on Schuco-Vac Suction Units should be completed a minimum of once each year.
2.
Suction Pump Preventative Maintenance:
Suction pump preventative maintenance includes the following:
Visually inspecting and cleaning external surfaces, controls, attachments and accessories.
Replacing the inlet filter.
Verification of the vacuum gauge.
Performance of ground resistant test and current leakage test for hospital grade plug units.
3.
Exterior Inspection:
Turn off the unit and disconnect from the AC or power source.
Inspect the exterior surfaces of the unit for broken or loose knobs. Check the power cord and plug for damage.
Replace the pump head filter, Part Number $641020, if damaged or worn.
Check the o-ring gasket, Part Number S633212, for wear and replace if necessary.