CPS BlackMax MD100HE Benutzerhandbuch - Seite 20

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte CPS BlackMax MD100HE herunter. CPS BlackMax MD100HE 20 Seiten.

APPENDIX H {ERROR INDICATOR SCREENS)
If certain malfunctions occur, MD50 and MD100 series will display error messages:
oP -Appears in display areas reserved for temperature readings when thermistor sensor
connected to that port is open circuited or not connected. Replace sensor with CPS model
number TMX3C.
SP -Appears in d isplay areas reserved for temperature readings when thermistor sensor
connected to that port is short circuited. Replace sensor with CPS model number TMX3C.
Err (Upper Left Display) - Appears in upper left display when T function is selected and
one or both thermistor sensors are defective as indicated by oP or SP indicators mentioned
above.
Err (Lower Left Display) -Appears in display areas reserved for temperature readings when
user tries to read saturation temperatures and calibration of pressure channels is invalid.
This defect requires manifold be returned to factory.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
BLACK
Digital Manifold Series
20
TestEquipmentDepot.com
®
73-154 REV B