CPS IAQPRO SmartAir Benutzerhandbuch - Seite 8

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte CPS IAQPRO SmartAir herunter. CPS IAQPRO SmartAir 18 Seiten. Professional indoor air quality meter
Auch für CPS IAQPRO SmartAir: Benutzerhandbuch (9 seiten), Handbuch für Schnellstart und Fehlersuche (8 seiten)

CPS IAQPRO SmartAir Benutzerhandbuch
QUICK START INSTRUCTIONS (CONT'D)
Step 3: Open the CPS Link App on your smart device and select the IAQPRO icon to connect unit.
Once connected, online guidance screens will appear prompting completion of Step 1 &
2 above. Swipe right to advance to the SET BASELINE screen. Touch the SET BASELINE
option to save tVOC baseline and complete setup.
CONDUCT AN INDOOR AIR QUALITY TEST
The tVOC sensor requires a stabilization period when powered ON (the light bar will flash Green/
Yellow during this process). The stabilization period can last between 5 to 60 minutes depending
on the combination of storage conditions and time (minutes, days, weeks) the unit was powered
OFF. Unit can be powered ON during travel between jobs to minimize jobsite stabilization time
using available automobile USB power connections or other readily available portable power
devices.
a. For more accurate readings and to generate an Indoor Air Quality Test Report, start a
30-minute test (default setting).
• Choose the "TESTS" icon (upper right of screen), then.
• Choose the "Air Quality" option (lower left of screen).
• Choose a "Standard Test" (30 minutes) or "Custom Test".
• If "Standard Test" selected, touch "Start Test" button to begin. A new screen will appear,
and the timer counts backwards from 30 minutes.
• To conduct a test shorter or longer than 30 minutes, select "Custom Test." Enter the
desired test period (minimum of 5 min to a maximum of 23 hrs, 59 mins, 59 seconds),
then press "Start Test".
b. The light bar on the SmartAir housing will display a
color indicating the range (see chart at right) within
which the highest reading for ONE of these values
(PM
, PM
, tVOC or CO
2.5
10
SmarAir is paired to your mobile device with the CPS
Link app, it will emit a a short blue flash once every
5 seconds
c. For more detailed information about any item being
measured, see the Information Section in the app.
8
) is found. Note: If the
2
GREEN
Good
YELLOW
Satisfactory
DARK YELLOW
Moderate
ORANGE
Poor
RED
Severe