3A HEALTH CARE HAPPYNEB III Handbuch - Seite 7

Blättern Sie online oder laden Sie pdf Handbuch für Medizinische Ausrüstung 3A HEALTH CARE HAPPYNEB III herunter. 3A HEALTH CARE HAPPYNEB III 17 Seiten.
Auch für 3A HEALTH CARE HAPPYNEB III: Gebrauchsanweisung (17 seiten)

3A HEALTH CARE HAPPYNEB III Handbuch
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations
as provided by disinfectant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.).
EN554/ISO11134.The sterilization packaging must conform to EN868/ISO11607 and be suitable for steam sterilization.
After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.

MAINTENANCE, CARE, AND SERVICE

Order all spare parts from your dealer or pharmacist.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles. Check
the filter continually for cleanliness and replace it if dirty, or after a maximum of 3 months use. Spare filters are provided with
the device. To replace the filter, open the air filter compartment 4 marked «Filter» and change the filter 11
MALFUNCTIONS AND ACTIONS TO TAKE
The device cannot be switched on
Ensure the power cord 2 is correctly plugged into the socket. Ensure the ON/OFF switch 3 is in the position «I». The device
has automatically switched off due to overheating. Wait until it has cooled down and try again.
The nebuliser functions poorly or not at all
Ensure the air hose 6 is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If ne-
cessary, replace with a new one. Ensure the nebuliser 5 is correctly assembled 12. Ensure the required medication has been
added.
GUARANTEE
This device is covered by a 3 year guarantee from the date of purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) conforming date of purchase or the receipt.The guarantee covers only the
compressor. The replaceable components like nebuliser, masks, mouthpiece, air hose, and filters are not included. Opening or
altering the device invalidates the guarantee.The guarantee does not cover damage caused by improper handling, accidents
or non compliance with the operating instructions.

TECHNICAL DATA

Piston electrocompressor with thermal cut
Nominal voltage: see data label
Dimensions: 310x130x220 mm
Weight: 2.0 Kg approx.
Temporary use: 30/30 min. I/O
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 250 kPa
Air flow: 14 Lt/min
Noise level: 56 dBA (EN13544-1)
Neb rate: 0.45 ml/min (con soluzione NaCl0.9%)
Operating pressure: 90 kPa (0.9bar)
Operating flow: 6 l/min
MMAD: 2.38 μm GSD: 2.84 Aerosol Output: 160 μl Output Rate: 100 μl/min
3A HEALTH CARE S.R.L.
Via Marziale Cerutti 90/F,
25017 Lonato (BS) - ITALY
0434
7