3A HEALTH CARE MAXIASPEED Gebrauchsanweisung - Seite 5

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Medizinische Ausrüstung 3A HEALTH CARE MAXIASPEED herunter. 3A HEALTH CARE MAXIASPEED 12 Seiten.

• In view of their length, the power cord and connection hose could constitute a strangulation hazard.
• The device is equipped with a safety fuse, easily inspected in the event of a fault, located on the back of the device in the panel
socket. Pull out the power supply plug before inspection.
• The antibacteria filter is a disposable element which must be replaced after every application.
• The device comes into contact with the patient via the single-use cannula (NOT supplied with the device): any aspiration
cannulas, purchased separately from the device, must comply with the requirements of the ISO 10993-1 standard on the bio-
compatibility of materials.
• Do not lift the device using the transportation handle in any circumstance.
• Before each use, follow the cleaning and disinfection operations as shown in the "CLEANING AND DISINFECTION OPERA-
TIONS" paragraph of this instruction manual.
Before each use, ensure that all accessories are perfectly clean according to the instructions in the "CLEANING AND DISINFECTION
OPERATIONS" section.
1. Connect the device as shown in figure 1 (page 2):
1.1 Connect the antibacterial filter (2.) to the "INLET" connector of the aspirator, ensuring that the side marked "IN" is facing
towards the connector. Do not use the aspirator without the antibacterial filter, because from a bacteriological point
of view, it becomes dangerous for the patient.
Connection
1.2 Connect one end of the short silicone tube to the hose connection of the antibacterial filter and the other end to the
"VACUUM" connection (4.1) of the vessel. The collection vessel is complete with overflow valve, lid and vessel in a clear mate-
rial (polycarbonate). The antibacterial filter also protects the suction circuit from possible contaminants sucked in during use.
1.3 Connect the long tube to the "PATIENT" connector (4.2) of the selected plug.
2. Operation of the telescopic handle: for space reasons the aspirator is supplied with the telescopic handle lowered; to
raise it, pull it up until the travel end hooks engage. To put the telescopic handle down again, press the travel end hooks on
the external sides of the telescopic tubes with the tool provided (Fig. 2) and push the handle downwards until it closes on the
aspirator casing.
3. Using the vacuum regulator (1.4) it is possible to set the desired vacuum value (bar). By turning the knob towards "+" a higher
vacuum is obtained, while turning it towards "-" a lower vacuum is obtained: these values can be read on the vacuum gauge
(1.2).
4. Start the device by switching the main switch to the "I" (ON) position (1.5).
Version with pedal switch: If the device is equipped with the pedal switch, to use it connect the plug on the cable to the
socket provided on the panel (1.10) on the back of the device and turn the switch (1.6) next to the main switch (1.5) to "I"
(ON), allowing the device to be switched on/off using the pedal switch.
5. After the application, switch of the device, remove the power supply cable from the current socket and perform the cleaning
operations as illustrated in the "CLEANING AND DISINFECTION OPERATIONS" paragraph.
6. There is a very capacious compartment (1.9) on the back of the device for putting the power supply cable back in after the ap-
plication, and if the device is equipped with the pedal switch, also for this latter.
INSTRUCTIONS FOR USE

FLOW DIRECTION

Vessel
IN
IN
10
Device
Connection