Available versions of the
electronic module
The electrical connection of all the devices is made through
a male fixed connector.
2
12-36 V DC
transistors,
filtered and
NPN and
regulated
PNP
energized via the
Bürkert transmitter
1 NPN
the flow sensor is
transistor
connected to
1 sinus
without
output
English
12
Type of fluid
Fluid group 1, art. 4 §1.c.ii
Fluid group 2, art. 4 §1.c.ii

Mechanical data

Part
Housing
Nut
Female connector type
2508 / screw / seal
Frame of the flow sensor
and paddle wheel
Axis and bearings
Gasket
Dimensions of device
please refer to the technical data sheets regarding the
type 8020 available at:
www.burkert.com
English
14
Hall, short
419 587
Hall, long
419 589
Hall Low
419 591
Power, short
Hall Low
419 593
Power, long
419 583
Coil, short
Coil, long
419 585
Conditions
DN ≤ 25
or PNxDN ≤ 2000
DN ≤ 200
or PN ≤ 10
or PNxDN ≤ 5000
Material
PE
PC
PA / Stainless steel / NBR
PVDF
Ceramic
FKM (EPDM optional)
6.

TECHNICAL DATA

Operating conditions
Ambient temperature -15 °C...+60 °C
Air humidity
< 80%, non condensated
Protection rating acc.
IP65, female connector wired,
to EN 60529
plugged and tightened
Conformity to standards and
directives
The applied standards, which verify conformity with the
EU Directives, can be found on the EU Type Examination
Certificate and/or the EU Declaration of Conformity (if
applicable).
• Article 4§1 of the Pressure Equipment Directive
2014/68/EU
The device can only be used in the following cases
(depending on the maximum pressure, the DN of the
pipe and the fluid):
Type of fluid
Fluid group 1, art. 4 §1.c.i DN ≤ 25
Fluid group 2, art. 4 §1.c.i
Fluid data
Fluid temperature
The fluid temperature may be
restricted by the fluid pressure:
Refer to the fluid temperature-
pressure dependency curves for
the device. SeeTable 1
• with fitting S020
• -15...+80 °C
in metal or PVDF
• with fitting S020
• 0...+80 °C
in PP
• with fitting S020
• 0...+50 °C
in PVC
Flow rate measuring range
• Hall and Hall Low
• 0,3...10 m/s
Power versions
• Sinus version
• 0,5...10 m/s
Measurement deviation
• with standard
• ±2,5% of the measured value*
K-factor
• with Teach-In
• +1% of the measured value*
Conditions
DN ≤ 32
or PNxDN ≤ 1000
English
English
13
13
15