ABB Endura AZ40 Handbuch - Seite 11

Blättern Sie online oder laden Sie pdf Handbuch für Industrielle Ausrüstung ABB Endura AZ40 herunter. ABB Endura AZ40 16 Seiten. Oxygen and carbon monoxide equivalent (coe) analyzer
Auch für ABB Endura AZ40: Gebrauchsanweisung (16 seiten), Betriebsanleitung (20 seiten)

ABB Endura AZ40 Handbuch
8.3 CO heater block thermocouple (T/C 1)
CAUTION – Minor injuries
The O
sensor must be removed to enable access
2
to COe heater block thermocouple T/C 1 (steps 1 to 3).
Referring to Fig. 8.3:
1. Place the sensor assembly on a clean flat surface with the
flange studs pointing down.
2. Use a small flat-bladed screwdriver to depress the spring
connectors at terminal numbers 8, 9 and 10 and
disconnect CO heater block thermocouple T/C 1 wires
from top terminal block connectors A.
sensor A from manifold block B using a
3. Unscrew O
2
7
/
in. spanner (wrench) and withdraw the sensor.
8
CAUTION – Minor injuries
Do not overtighten when refitting the O
1
2
3
4
8
O2 HTR WHT+
Fig. 8.3 Removing the O
sensor
2
sensor.
2
5
6
7
8
8
9 10
9 10 11 12 13 14 15 16 17 18 19 20
10
9
O2 SEN BLK
O2 HTR WHT–
Endura AZ40 | Sensor assembly | Heater and thermocouple kit | INS/ANAINST/008-EN Rev. B
11