ABB Endura AZ40 Handbuch - Seite 8

Blättern Sie online oder laden Sie pdf Handbuch für Industrielle Ausrüstung ABB Endura AZ40 herunter. ABB Endura AZ40 16 Seiten. Oxygen and carbon monoxide equivalent (coe) analyzer
Auch für ABB Endura AZ40: Gebrauchsanweisung (16 seiten), Betriebsanleitung (20 seiten)

ABB Endura AZ40 Handbuch
7 Removing / Refitting sensor covers
7.1 Removing sensor covers
Referring to Fig. 7.1:
1. Unscrew 4 (captive) cover retaining screws A using a
medium flat-bladed screwdriver. and remove sensor cover
B.
2. Unscrew 2 (captive) cover retaining screws C using a
medium flat-bladed screwdriver and remove sensor
terminal cover D.
Retain covers B and D for re-use.
IMPORTANT (NOTE)
Check the seals fitted to each cover. If they need
replacing or seals have bonded to the mating
surfaces of the sensor assembly, replace them –
see Instruction INS/ANAINST/10-EN.
Fig. 7.1 Removing sensor covers
8
INS/ANAINST/008-EN Rev. B | Endura AZ40 | Sensor assembly | Heater and thermocouple kit
7.2 Refitting sensor covers
Referring to Fig. 7.1:
1. Apply a light coating of anti-seize compound (suitable
for temperatures up to 200 °C [392 °F]) to the threads
of sensor cover mounting screws A and C.
2. Refit the sensor cover B and sensor terminal cover D in
the reverse order of removal – refer to Section 7.1.
3. Prepare the sensor assembly for operation by reversing
the procedures in Sections 5 and 4 (pages 5 and 4).
4. Refer to Operating instruction OI/AZ40-EN to restore the
analyzer to full operational condition.