P6 - Bluetooth App Code
(Useful only when Bluetooth App is
Purchased)
This setting will enter Bluetooth App Authorization Code in the machine for
Android Bluetooth App and Meter Paring.
a) In Setting Mode select "P6" and press "OK" to enter code.
b) Press "UP" or "DOWN" to enter values, Press "LEFT" or "RIGHT" to
shift between variables.
c) Press "OK" to Confirm and "ESC" and Exit settings.
P7 - Restore Factory Setting
This setting will restore Meter to Factory Default Values.
a) In Setting Mode select "P7" and press "OK" Machine will display
"8888"
b) Press "OK" to Confirm and "ESC" and Exit settings.
P8 - Manual Temperature Compensation
(Default Value = 25 Deg C)
This setting will enter the manual temperature compensation value,
applicable only if Temperature Electrode is not connected.
a) In Setting Mode select "P8" and press"OK" to enter value.
b) Press "UP" or "DOWN" to enter values, Press "LEFT" or "RIGHT" to
shift between variable.
c) Press "OK" to Confirm and "ESC" and Exit settings.
P9 - Bluetooth Operation Mode
This setting will select Bluetooth Operation Mode
APP Mode - Use Android App to connect Meter
Bluetooth Mode - Use to print Meter reading on Bluetooth Printer
a) In Setting Mode select "P9" and press "OK" to select APP or
Bluetooth.
b) Press "UP" or "DOWN" to select Mode and press "OK" to confirm.
c) Press "ESC" and Exit settings.
P10 - Electrode Coefficient
(K) Value setting
This setting will enter Conductivity Electrode (K=1.0) value as mentioned on
the electrode.
a) In Setting Mode select "P10" and press "OK" to select value.
10
b) Press "UP" or "DOWN" to select value and press "OK" to confirm.
c) Press "ESC" and Exit settings.
P11 – Electrode Coefficient (K) Range setting
This setting will select Conductivity Electrode (K=0.1/1.0/10) range as
mentioned on the electrode.
a) In Setting Mode select "P11" and press "OK" to select range.
b) Press "UP" or "DOWN" to select range and press"OK" to confirm.
c) Press "ESC" and Exit settings.
P12 - Standard Solution Selection
This setting will select Standard Solution range as mentioned on the
electrode.
a) In Setting Mode select "P12" and press "OK" to select range.
b) Press "UP" or "DOWN" to select range and press "OK" to confirm.
c) Press "ESC" and Exit settings.
P13 - Temperature Compensation Coefficien
(Default = 2.00 % /Deg C)
This setting will enter Temperature Compensation Coefficient for Automatic
Temperature Compensation.
a) In Setting Mode select "P13" and press "OK" to enter value.
b) Press "UP" or "DOWN" to enter values, Press "LEFT" or "RIGHT" to
shift between variable.
c) Press "OK" to Confirm and "ESC" and Exit settings.
P14 - Datum Temperature Setting
This setting will select Datum Temperature (18 / 20 / 25) range as mentioned
on the electrode.
a) In Setting Mode select "P11" and press "OK" to select range (18 / 20
/ 25).
b) Press "UP" or "DOWN" to select range and press "OK" to confirm.
c) Press "ESC" and Exit settings.
(USA/NIST/CHINA)