PROCEDURE ONE: Use of GPS™ III Mini Separator to prepare platelet-rich
plasma (PRP)(Figure 1)
1. DRAW: Draw 3ml ACD-A into 30ml syringe. Attach to 18-gauge apheresis
needle and prime with ACD-A. Slowly draw 27ml of the patient's blood into
the 30ml syringe primed with ACD-A. Gently, but thoroughly, mix the whole
blood and ACD-A upon collection to prevent coagulation.
2. LOAD: ENSURE BLOOD FROM ONLY ONE PATIENT IS PROCESSED
PER SPIN, and that the GPS™ III Mini Separator remains upright.
Unscrew cap connected on center port #1. Remove and discard cap and
green packaging post. Slowly load blood filled 30ml syringe (3ml ACD-A
mixed with 27ml of the patient's whole blood) into center port #1. Unscrew
and discard protective inner piece from white cap tethered to port #1.
Screw white cap onto port #1. Place GPS™ III Mini Separator filled with
anticoagulated blood in a Biomet Biologics™ centrifuge.
3. BALANCE: Fill blue GPS™ counterbalance tube (800-0505) with 30ml of
sterile saline/water (equal to amount of whole blood plus ACD-A dispensed
in the GPS™ III Mini Separator). Place filled counterbalance directly
opposite from the GPS™ III Mini Separator in the centrifuge.
4. SPIN: Close centrifuge lid. Set RPM to 3.2 (x1000) and the time to 15
minutes. Press the start button. Once spin is complete, open the centrifuge
and remove the GPS™ III Mini Separator.
5. EXTRACT PPP: Unscrew and save yellow cap on port #2. Connect 30ml
syringe to port #2, invert GPS™ III Mini Separator, and extract platelet-poor
plasma (PPP). Remove 30ml syringe from port #2, cap with a sterile syringe
cap and set aside. Replace yellow cap on port #2.
6. SUSPEND PRP: Holding GPS™ III Mini Separator in the upright position,
unscrew red cap on port #3. Attach sterile 10ml syringe to port #3. Extract
1ml of PRP into the 10ml syringe. Leave the syringe attached. Shake
GPS™ III Mini Separator gently for 30 seconds.
7. EXTRACT PRP: Immediately after suspending the platelets, extract the
remaining PRP into the attached 10ml syringe. Remove 10ml syringe form
port #3 and cap with a sterile syringe cap.
PROCEDURE TWO: Use the Clotalyst™ Separator to prepare the autologous
serum (Figure 2)
1. TURN ON: The Clotalyst™ Heater to achieve the appropriate operating
temperature 25°C (77°F).
2. DRAW: Draw 1ml ACD-A into a sterile 10ml syringe. Attach to the 18-gauge
apheresis needle. Slowly draw 11ml of the patient's blood into the 10ml
syringe primed with ACD-A. Gently, but thoroughly mix the whole blood
and ACD-A upon collection to prevent coagulation.
NOTE: The 10ml syringe provided has graduation markings to 12ml.
3. LOAD Clotalyst™ REAGENT: Ensure the Clotalyst™ Separator remains
upright. Unscrew and discard blue cap on port #2, attach Clotalyst™
Reagent syringe and completely transfer contents (4 ml) through port #2.
Remove and discard reagent syringe. Unscrew and discard protective inner
piece from white cap tethered to port #2. Screw white cap onto port #2.
NOTE: Dislodging duckbill valves attached to ports #2 and #3 of the
Clotalyst™ Separator during steps 3 and 4 can occur. This is NOT
detrimental to the procedure.
4. LOAD BLOOD: ENSURE BLOOD FROM ONLY ONE PATIENT IS
PROCESSED Unscrew and discard red cap on port #3. Attach and slowly
transfer contents of anticoagulated blood-filled syringe (1ml ACD-A mixed
with 11ml of patient's whole blood) through port #3. Unscrew and discard
protective inner piece from white cap tethered to port #3. Screw white cap
onto port #3.
5. MIX: Without producing foam gently invert Clotalyst™ Separator 12 times
to thoroughly mix the anticoagulated blood with the Clotalyst™ reagent
and glass beads.
6. INCUBATE: Holding the Clotalyst™ Separator on its side ensures the glass
beads are distributed evenly. Place the Clotalyst™ Separator in the heater
and begin the incubation cycle (25 minutes at 25°C).
7. SHAKE: Remove Clotalyst™ Separator from the heater and shake
vigorously for 5 second, thus dislodging and braking up any coagulum
The information contained in this package insert was current on the date this brochure was printed. However, the package insert may have been revised after that date. To obtain a current package
insert, please contact Biomet at the contact information provided herein.
that may be present in the separator. Place the Clotalyst™ Separator in a
Biomet Biologics™ centrifuge.
8. BALANCE: Place the orange Clotalyst™ counterbalance (800-0760)
directly opposite from the Clotalyst™ Separator in the centrifuge.
9. SPIN: Close centrifuge lid. Set RPM to 3.2 (x 1,000) and the time to 5
minutes. Press the start button. Once the spin is completed, open the
centrifuge.
10. EXTRACT SERUM: Gently remove the Clotalyst™ Separator without
disturbing the stratified layers. Holding Clotalyst™ Separator upright
unscrew and remove the yellow cap on center port #1. Attach a sterile
10ml syringe to center port # 1. To maximize serum volume output gently
tilt Clotalyst™ Separator toward port #3 while slowly extracting 5-6ml
serum Remove the 10ml syringe from center port #1 and attach sterile
syringe cap
NOTE: If the extraction tube becomes plugged prior to achieving
5-6ml of output, the Clotalyst™ Separator can be placed back in the
centrifuge for one minute at 3200 RPMs to dislodge the clot, allowing
the remaining serum to be extracted.
11. STORAGE: Store serum at 18-26° C (64.4-78.8°F) for up to 4 hours prior to
use. Discard remaining serum if not used within the provided time frames.
Port # 1
Port # 2
Duckbill
Extraction Tube
Figure 2
Clotalyst™ Separator
Caution: Federal law (USA) limits this device to sale, distribution, or use by or on
the order of a physician.
Comments regarding this device can be directed to Attn: Regulatory Dept.,
Biomet, Inc., P.O. Box 587, Warsaw, IN 46581 USA, Fax: 574-372-3968.
All trademarks herein are the property of Biomet, Inc., or its subsidiaries unless
otherwise indicated.
Manufacturer
Date of Manufacture
Do Not Reuse
Consult Accompanying Documents
Sterilized using Ethylene Oxide
Sterilized using Irradiation
Sterile
Sterilized using Aseptic Technique
Sterilized using Steam or Dry Heat
Expiry Date
U U
WEEE Device
Catalogue Number
Lot Number
Filter
Port # 3
Tethered Cap
Protective Inner
Pieces/syringe Tip
Glass Beads