BioNano Genomics Saphyr System Handbuch zur Systemsicherheit - Seite 6

Blättern Sie online oder laden Sie pdf Handbuch zur Systemsicherheit für Laborausrüstung BioNano Genomics Saphyr System herunter. BioNano Genomics Saphyr System 9 Seiten.

Laser Safety Warnings

The Saphyr system has a laser system classification of a Class 1 product, which uses embedded Class 3B and
one Class 4 laser. The Saphyr System is certified to the IEC 60825-1:2014 standards.
The nominal wavelength outputs include the following optical power:
500 mW (532 nm)
140 mW (637 nm, 2 each)
400 mW (488 nm)
90 mW (785 nm)
A Class 1 laser product is safe under all conditions of normal use.
A Class 4 laser product is hazardous if the eye is directly exposed to the laser. However, the diffuse reflections
are not harmful. Protective eye wear is required when a person is viewing a class 3B laser beam. All Class 4
lasers must be equipped with a key switch and a safety interlock.

Preventing Radiation Exposure

Do not open the front panel of the instrument because you could come into contact with a Class 4 laser.
If you operate the instrument with any panels removed, you risk radiation exposure from the direct or
reflected laser light.
If you use controls or make adjustments that are not specified in The Saphyr Safety Guide, you can be
exposed to hazardous radiation.

Safety Labeling

Figure 1: Class 4 Laser Warning
30253 Rev A, Saphyr System Safety Guide for Saphyr P/N 60325
Page 6 of 9