Operating the NESS H200 Wireless System
RF Communication
Indicator
Control Unit
Status Indicator
Intensity Level
Turn on the system. Press the
1
Test the position of the Orthosis. Press and hold the
2
test the extensor muscles. Release the button and press and hold again to test the flexor
muscles. Repeat as needed.
Select Program 1 or Program 2. Press the Program 1 or Program 2 selection button on
3
the side of the Control Unit. The selected program indicator will be GREEN:
If necessary, adjust the stimulation intensity level. Press the
4
button to increase/decrease stimulation intensity in the Orthosis.
Start a program. Press the
5
the Orthosis to turn on stimulation. To pause/resume a program, press the trigger button
on the Control Unit or the trigger button on the Orthosis.
Note: Neuroprosthesis Programs cannot be paused. Pressing the trigger button when
stimulation is on initiates the next phase of the program. To turn off stimulation, press the
on/off button.
H200 Wireless Control Unit Displays
On/Off Button Flashes Green
Trigger Button Flashes Yellow Rapidly
Trigger Button is a Steady Yellow
Program 1 Indicator is Green
Program 2 Indicator is Green
Control Unit Status Indicator is a Steady Red
RF Communication Indicator Flashes Red
Rx Only
(US Only)
For a complete list of all contraindications, warnings, adverse reactions, precautions, and care instructions please
review the User's Guide found at www.bioness.com.
© 2015
NESS
, NESS H200
Wireless, Bioness, the Bioness Logo
®
®
1
On/Off Button
Program 1 Indicator
Program 2 Indicator
on/off button.
trigger button on the Control Unit or the trigger button on
and LiveOn
are registered trademarks of Bioness Inc. in the United States or other countries. | www.bioness.com
®
®
Trigger Button
5
4
stimulation test button once to
System is On
Stimulation is On
Stimulation is Paused
Program 1 Selected
Program 2 Selected
Control Unit Malfunction: Charging Error; Registration
Error; Software Error
RF Communication Error
Intensity Adjustment Buttons
Mute Button
Program 1
Selection Button
3
Program 2
Selection Button
Stimulation Test
2
Button
or
plus or
.
minus