Alpha Moisture Systems PSS Benutzerhandbuch - Seite 9
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Probenehmer Alpha Moisture Systems PSS herunter. Alpha Moisture Systems PSS 17 Seiten.
PSS - Portable Sample System - User Manual
2.2 Continued.
Fig 4.
Pressure Regulator fully closed
anti-clockwise
2.3 Before connecting a sample line to the Isolation valve at the gas source, first check that
the gas to be sampled is free from excessive moisture and/or particle contaminants by
placing a white filter paper (or similar) approximately 1 cm from the outlet and very
slowly and carefully opening the Isolation Valve to achieve a steady flow of gas/air onto
the white filter paper for approximately 1 minute. See Fig 6.
Fig 6.
Isolation Valve Test
If contaminants are seen on the white filter paper, a filter unit must be installed at the
Isolation Valve outlet and before the sample hose.
When the filter unit is fitted, repeat the above procedure with the white filter cloth.
For Technical Help Contact Alpha Moisture Systems.
Fig 5.
Needle Valve fully closed clockwise
Sample Point Isolation Valves, as
pictured in Fig 7. are not supplied with
the PSS - Please contact your local
distributor on how to order.
9
NOTE