Axis Electronics NANO-200G2 Benutzerhandbuch - Seite 15

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Skalen Axis Electronics NANO-200G2 herunter. Axis Electronics NANO-200G2 17 Seiten. Electronic weighing scale

Axis Electronics NANO-200G2 Benutzerhandbuch
Pause for 2 seconds at the previous step
and press Zero/Enter key. The display
4
shows "LoAd" momentarily and shows the
internal counts.
Put a known weight on the weighing pan
(Ex.: 200g). Press the Enter key. The
5
display will now show all "000" with the
units digit blinking.
If you put 200g weight, Enter "200".
Change the blinking digit by pressing the
6
mode button. Shift the changed digit to
left by pressing the "P" button once.
Press the Enter key once again. The display
7
will show "......" for some time and then
show the weight
With the weight still on the pan, the
display will now show the weight that is
8
calibrated. Check calibration to ensure the
scale is reading correct.
Turn of the machine. Replace the CAL
jumpers back to the original position.
9
Tighten the screws and close the covers of
the machine as it was in the original
condition.

ERROR CODES

ERROR – DISPLAY
Error
Zero not in range
15
NANO Series Precision Balance
Remove weight from pan
and restart the machine.
If the problem persists,
the machine may have
been overloaded.
Solution