Olympus BX3-25ND6 Anleitung Handbuch - Seite 29

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus BX3-25ND6 herunter. Olympus BX3-25ND6 36 Seiten.

Olympus BX3-25ND6 Anleitung Handbuch
Fig. 16
3
Fig. 17
Attaching the Mercury Burner and
3
1
Lamp Housing5
1. Loosen the socket clamping screw @ using the Allen screwdriver.
2. Hold the upper section of lamp housing and pull it upward to remove
the socket section.
2
To prevent malfunctions, do not hold the lamp housing by the
CAUTION
centering knobs 2.
3. Place the socket section upside down as shown in Fig. 17.
} The lamp housing is equipped with the holder for transportation in the
factory shipment condition, or with an old burner when the burner is
replaced. Remove the holder or old burner by loosening the two burner
holding screws 3
4. Attach the + (positive) pole of the mercury burner | to the fixed mount
on the upper side, then the − (negative) pole to the mount on the lower
side.
· Be sure to use the USH-103OL (mfd, by OLYMPUS) burner.
CAUTION
4
· Be careful and avoid leaving fingerprints or contaminants on
the mercury burner. Otherwise, there is a danger of explosion
due to distortion of glass caused by the stains. If the burner
is contaminated, clean it by wiping gently with gauze slightly
moistened with absolute alcohol.
5. Attach the socket section with burner to the original position and tighten
the clamping screw @.
· Align the external edges of the lamp housing with those on
the socket section, and push the lamp housing straight down-
CAUTION
ward.
· Attach the lamp housing on the lamp housing installation po-
sition of the microscope system so that the heat radiating fins
face upward. To prevent a fire hazard, reserve ample spaces
above, below and on the rear of the lamp housing.
· Do not light the mercury burner while it is not mounted on the
microscope because the UV rays in its light are harmful to your
eyes.
· The UV rays in the light of the mercury burner may damage the
specimen if this is sensitive to UV rays.
}
CAUTION
6. Connect the connector of the lamp house to the power unit U-RFL-T.
7. Connect the power cord to the power unit, and insert the power plug into
the outlet.
Burner Service Life
USH-103OL : 300 hours
This value assumes light cycles composed of 2 hours of
lighting and 30 minutes of extinction. Do not turn it on and
off at a shorter cycle than the above, for this will shorten
the service life of the burner.
After replacing the burner, reset the hour counter of
the power supply unit to "0.0."
(Figs. 16 & 17)
26