5-3 Conoscopic Observation
Fig. 32
}Use 20X to 100X objectives.
1. Engage the analyzer and adjust for extinction position.
2. Swing the condenser top lens into the light path.
3. When using the polarizing intermediate attachment U-CPA, push in the
Bertrand lens knob to engage the Bertrand lens into the light path.
4. Engage an objective of your choice (20X to 100X).
5. Open the aperture iris diaphragm.
6. Slide the focusing knob on the U-CPA to focus on the conoscopic image.
}When the polarizing intermediate attachment U-CPA is not used,
conoscopic observation is possible by removing an eyepiece from the
observation tube and looking at the rear focal plane of the objective
directly.
}To obtain good image contrast, place an interference filter (45IF546) in
the filter holder on the light exit of the microscope.
}If the periphery of the conoscopic image is dark, move the condenser
vertically to find the position where the periphery is brightest.
BX51/52-P,BX41-P