Olympus BX41-P Anleitung Handbuch - Seite 22
Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus BX41-P herunter. Olympus BX41-P 36 Seiten. Polarizing microscope
Auch für Olympus BX41-P: Anleitung Handbuch (36 seiten)
Fig. 24
@
Fig. 25
19
²
Cross lines
@
Conoscopic image
Note: During conoscopic observation, the center of the conoscopic image
@ and the intersection of the cross lines ² may not coincide.
However, employment of the universal infinity optical design ensures
that this disagreement will have no measurable effect during actual
observation.
However, if the disagreement is distracting, adjust in the same manner
as when using the U-CPA (for photomicrography), described in the
following paragraph.
When Using Intermediate Attachment for conoscopic and orthoscopic
observation (U-CPA)
· Photomicrography
1. Using the tip of a mechanical pencial, etc., push out, in the direction of
the objective hole, the black rubber plugs inserted into the centering
holes of the primary hole where the 20X or 10X objective is inserted.
2. Pull out the Bertrand lens knob @ to the OUT (¦) position. (Fig. 25)
3. Adjust the position of the intermediate tube (Refer to page 7.)
4. Center the condenser. (For details, see instructions of each microscope.)
5. Push in the Bertrand lens knob @ to the IN ({) position for conoscopic
observation.