Olympus BX51-P Anleitung Handbuch - Seite 19
Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus BX51-P herunter. Olympus BX51-P 36 Seiten. Polarizing microscope
Auch für Olympus BX51-P: Informationen (6 seiten), Anleitung Handbuch (36 seiten)
POLARIZED LIGHT OBSERVATION
5-1 Adjustments Before Observation
In polarized light microscopy ultimate performance is not obtainable unless optical adjustments are made correctly.
Always performed the following adjustments before observation. Remove the quarter wave plate and sensitive tint plate
from the light path.
Fig. 19
1
Adjusting the Optical Axis
When Using the Intermediate Attachment for AN360P (U-OPA) (Figs. 19 & 20)
³
# Make sure the 10X objective is inserted into the primary hole on
²
the centerable revolving nosepiece.
1. Fully insert the alignment polarizer @, provided with the U-OPA into
the test plate adapter (U-TAD). Tighten its clamping knob. (Fig. 19)
@
2. Remove the condenser.
3. Insert the rotatable analyzer (U-AN360P-2) ² into the analyzer slot on
the intermediate attachment. Engage the analyzer, loosen the clamping
screw and set the analyzer's rotation dial ³ to 0°. (Fig. 19)
BX51/52-P,BX41-P
16