Attaching the Stage (U-SIC4, U-SV)
,Rg.,,
1. FuUy loosen the clamping screw © at the rear of the stage (at the
front n the case of U-SV).
2.
then tighten the clamping screw.
fig. 1
Rg.2
Mounting the Condenser
)FIa. a
1. Turn the coarse adjustment knob © to raise the stage to its upper
limit
lAssuming there are no objectives mounted on the nosepiece yet.)
2. Tumiheco^r^heiBht
holder to its tower Emit
3. Loosen the condenser clamping screw.
4. Position the condenser with the scale markings in front and insert it
into the substage fork as far as it wB go.
5. Tighten the condenser clamping screw, then raise the condenser to
its upper limit
in tha substage fork.
™Ullin9 *• U'SC «•**•<"« «chrom«le condenser or the
U-UCO universal condenser, swing the top lens out of the way
before inserting the condenser.