Olympus CX41 Anleitung Handbuch - Seite 25

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus CX41 herunter. Olympus CX41 32 Seiten. System microscope
Auch für Olympus CX41: Übersicht (7 seiten), Anleitung (32 seiten)

Olympus CX41 Anleitung Handbuch

7-2 Detailed Assembly Procedure

Fig. 25
Fig. 26
1
Mounting the Bulb (Replacement of Bulb)
1. Turn the microscope frame on its side and pull the lamp housing knob
@ on the underside of the base to open the lamp housing cover.
2. Holding the halogen bulb ² contained in the polyethylene bag to
avoid leaving fingerprints on the bulb, fully insert the contact pins into
the bulb socket ³ When properly seated, pull off the polyethylene bag.
<Applicable bulb>
6 V, 30 W halogen bulb: 6V30WHAL (Philips 5761)
Always use the designated bulb. Use of improper bulb may result
in fire.
Do not touch the bulb with bare hands. If fingerprints are
accidentally left on the bulb, wipe the bulb with a soft, lint-free
cloth moistened with alcohol. Using a contaminated bulb will
shorten the service life of the bulb.
3. With the lamp housing knob still pulled out, close the lamp housing
cover. Then push in the lamp housing knob to lock the cover.
#The lamp housing cover cannot be closed if the knob is pushed in
before you attempt to close the cover.
Caution for Bulb Replacement During Use or Right After Use
Whenever you replace the bulb during use or right after use, first
move the main switch to "
from the wall outlet, and allow the bulb and parts around the bulb
to cool before touching.
#If the bulb burns out during an observation and needs to be
replaced, remove eyepieces, specimen, filter and other objects
likely to fall off, before tilting the microscope frame to replace the
bulb.
5
Mounting the Condenser
1. Turn the coarse adjustment knob @ to raise the stage to its highest
position.
2. Turn the condenser height adjustment knob ² to lower the condenser
holder to the position where the clamping screw ³ can be turned.
3. Insert the condenser | all the way into the mounting hole ƒ and
clamp by tightening the clamping screw ³. Positioning the condenser
with the diaphragm lever ... at the front facilitates operation.
4. Turn the condenser height adjustment knob ² to raise the condenser
holder to its highest position.
(Fig. 25)
" (OFF), disconnect the power cord
(Fig. 26)
CX41
22