Olympus KHC Anleitung Handbuch - Seite 10
Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Mikroskop Olympus KHC herunter. Olympus KHC 12 Seiten.
VIII.
Major troubles with microscope, their possible
causes and applicable steps to be taken.
Troubles
I
Possible causes
1
Steps to be taken
Slipping down of the
Loosened coarse
See Smoothening of Coarse
stage by itself
adjustment knob
Adjustment knob and pro-
perly adjust it.
Coarse Adjustment
1.
Axis of Coarse Adjust-
1. Take reverse steps of
knob too heavy
ment knob overly
the above.
tightened
2.
Hardened grease on the
2.
Notify the
sliding surface.
manufacturer.
-
Fine Adjustment knob
Hardened grease on the
Notify the manufacturer.
too heavy
sliding surface
Dust/Oirt on objective
1. Eyepiece tubes are left
Wipe off the top lens of the
without eyepieces or
objective with clean gauze;
cap.
do not touch inner lenses
2.
Fingerprints left on the
but notify the manufacturer
lens;
Lens hitting the speci-
men;
or cargille oil left on
the oil-Immersion lens.
Dust/Dirt on eyepiece
1.
Careless storage.
1. Be sure to store the
eyepieces after use in
the provided container
or in a desiccator
2.
Fingerprints left on the
2.
Wipe off with clean
lens surfece
gauze. Do not try to
disassemble the
eyepiece. Any change in
arrangement of lenses
can deteriorate the
entire opical efficiency.
Scratches on the lens
Lens was wiped with a piece
Immediately notify the
of cloth or leather while
manufacturer. The lens will
dust or filth is still left on.
be replaced at the actual
cost.
-
7 -