Olympus MIC-D Reparatur-Handbuch - Seite 28

Blättern Sie online oder laden Sie pdf Reparatur-Handbuch für Mikroskop Olympus MIC-D herunter. Olympus MIC-D 43 Seiten. Digital microscope
Auch für Olympus MIC-D: Schnellstart-Handbuch (28 seiten)

Olympus MIC-D Reparatur-Handbuch
MIC-D
MIC-D
(4)Adjustment
(4)Adjustment
-1-Set the cross-specimen on stage.
-1-Set the cross-specimen on stage.
-2-Connect the USB cable with PC.
-2-Connect the USB cable with PC.
-3-Set the zoom of MIC-D to highest magnification position.
-3-Set the zoom of MIC-D to highest magnification position.
-4-Focus the cross-specimen by focusing knob.
-4-Focus the cross-specimen by focusing knob.
-5-Set the zoom of MIC-D to lowest magnification position and Check the parfocality.
-5-Set the zoom of MIC-D to lowest magnification position and Check the parfocality.
highest magnification position
highest magnification position
   
When parfocality is no good
When parfocality is no good
-6-Remove the rubber feet and bottm plate. ACU3X4SA(2pcs.)
-6-Remove the rubber feet and bottm plate. ACU3X4SA(2pcs.)
-7-Remove the C-MOS ass'y(CUKSK3X4SA:4pcs.) and light intensity adjustment board.(CUKSK3X6SA:2pcs.)
-7-Remove the C-MOS ass'y(CUKSK3X4SA:4pcs.) and light intensity adjustment board.(CUKSK3X6SA:2pcs.)
-8-Assemble the MIC-DKC02 to base ass'y. Hold the parfocality lens by MIC-DKC02.
-8-Assemble the MIC-DKC02 to base ass'y. Hold the parfocality lens by MIC-DKC02.
Light intensity adjustment board
Light intensity adjustment board
D.ADJUSTMENT PROCEDURE
D.ADJUSTMENT
C-MOS ass'y
C-MOS ass'y
D-3
D-3
PROCEDURE
lowest magnification position
lowest magnification position
MIC-DKC02
MIC-DKC02
Parfocality lens
Parfocality lens