3

Operation

3-1 Using the stray light prevention cover

b
a

3-2 Diopter adjustment

b
1
0
10
20
30
40
50
60
70
1
Cover the eyepieces a which are not used by the assistant observers
with the provided stray light prevention covers b .
1

Diopter adjustment of main observer

When the eyepiece is equipped with the eyepiece micrometer
1
While looking through the eyepiece equipped with the eyepiece
a
micrometer, rotate the diopter adjustment ring b to adjust so that
the scales or lines of the eyepiece micrometer in the field of view are
clearly visible. When you rotate the diopter adjustment ring b , keep
pressing the lower part a of the eyepiece.
2
Place the specimen.
3
Engage the 10X objective in the light path. While looking through the
eyepiece equipped with the eyepiece micrometer, rotate the coarse/
fine focusing knobs to bring the specimen into focus.
4
Rotate the diopter adjustment ring b of the eyepiece which is not
equipped with the eyepiece micrometer to bring the specimen into
focus.
80
90 100
11