PREVENTIVE INSPECTION SHEET FOR ILLUMINATION DEVICES

{We recommend performing "Preventive Inspections" periodically (every time you replace lamps and at least once every 6
months).
{The table below identifies the check items to be observed. Put (X) if not applicable or ( ) if applicable.
{If there are any check marks ( ) noted, immediately stop use of the product and seek service or replacement for the
illumination device(s) for prevention.
{If you detect an abnormality other than that listed below with your illumination device or other Olympus product, request
inspection from your Olympus distributor.
1. More than 8 years have passed since original purchase or exceeds 20,000
hours of use.
2. Lamp does not light sometimes even through the power is on.
(Except discharge burners
3. Light flickers when you move a lamp cable or illumination devices.
4. Lamp cable is unusually hot to the touch.
5. Burning or smoke odor.
6. Light still flickers after new lamp replacement.
(Except discharge burners
7. Sings of deformation, backlash, or looseness, etc. when you assemble/
disassemble the illumination device.
(I.e. it is hard to open/ close the lid during lamp-replacement.)
8. Connection terminals or a lamp installation terminal have become discolored
or tarnished on the right or left side.
(Except discharge burners
9. Illumination device/ housing has become deformed, cracked or tarnished in
any way.
10. Lamp cables or wiring parts have become deformed, cracked or tarnished in
any way.
11. Frequent repairs to similar devices put into use at the same time as the unit
being checked.
*1
Discharge burners: Mercury burner / xenon burner / metal halide burner
13
If you have any questions, please contact your Olympus distributor.
Check items