Olympus BX3-SSU Anleitung Handbuch - Seite 8

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Medizinische Ausrüstung Olympus BX3-SSU herunter. Olympus BX3-SSU 20 Seiten. Scanning stage with ultrasonic
Auch für Olympus BX3-SSU: Anleitung Handbuch (24 seiten)

Olympus BX3-SSU Anleitung Handbuch
XY controller (Provided with BX3-SSU)
* The factory settings of the stage displacement per turn are 10 mm for the coarse adjustment and 0.2 mm for the fine
adjustment.
Auto Compensation of the Stage
This stage is equipped with the Auto Compensation for the stage operation parameters to provide you the best operational
feeling to fit to your use environment.
If the stage operation parameters are displaced against your use environment, you may not obtain the best operational
feeling, such that the stage does not move smoothly, etc. In that case, perform the Auto Compensation of the stage
according to the following procedures to reset the stage operation parameters to the optimum value.
Before starting the auto compensation, install the stage on the BX63 and make sure nothing, including a
CAUTION
specimen, is placed on the stage.
1
2
5 5
Coarse/fine selector button
Pushed in: Coarse adjustment.
Out position: Fine adjustment.
1. Open the [Menu] display by controlling the touch panel.
2. Tap the System Setting button @ to open the [System Setting] display.
3. Tap the [ECO, Others] tab 2 and then tap the Auto Compensation
4. When the auto compensation confirmation display appears, tap [OK] to
CAUTION
5. When auto compensation completes, the touch panel screen shows the
CAUTION
3
button 3.
start auto compensation.
The stage moves automatically during the auto compensation
(for a few minutes). Do not touch the stage during the
movement.
completion message.
If the foreign substance is entered in the stage driving part,
the status may not be recovered by performing the Auto
Compensation. In this case, turn the power of BX3-SSU Control
Box to "
" (OFF), and perform the Auto Compensation again
after moving the stage from front to back and from side to side
several times with a hand.
Y-axis knob*
X-axis knob*