Olympus ESG-300 Kurzreferenz-Handbuch - Seite 9
Blättern Sie online oder laden Sie pdf Kurzreferenz-Handbuch für Medizinische Ausrüstung Olympus ESG-300 herunter. Olympus ESG-300 10 Seiten. Electrosurgical generator
Auch für Olympus ESG-300: Kurzreferenz-Handbuch (3 seiten), Kurzreferenz-Handbuch (4 seiten)
ESG-300
( Electrosurgical Generator)
Quick Reference Guide
ESG-300 Operation Settings
Argon Plasma Coagulation (APC)
Procedure
Diffuse bleeding in the
right colon/duodenum
Diffuse bleeding in the
remaining colon and rectum
GAVE/radioproctitis
Angiodysplasia
Tumor Reduction
Barrett's Esophagus
Pulmonary Procedures
Procedure
Less extended lesion,
bleeding
Polypoid Lesion
(intraluminal mass)
Web-like Stenosis
Tissue Ablation
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate
mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions,
Olympus cannot accept liability for clinical risks arising from the use of these example settings.
Instrument or Technique
Activate MAPC probe within a
2-5 mm distance from tissue
Activate MAPC probe within a
2-5 mm distance from tissue
Activate MAPC probe within a
2-5 mm distance from tissue
Activate MAPC probe within a
2-5 mm distance from tissue
Activate MAPC probe within a
2-5 mm distance from tissue
Activate MAPC probe within a
2-5 mm distance from tissue
Instrument or Technique
CD-6C-1
SD-7C-1
KD-31C-1
Activate MAPC probe within a
2-5 mm distance from tissue
Mode
Power
SmartArgon
120
PulsedArgon Fast
30
PulsedArgon Fast
30
PulsedArgon Fast
30
ForcedArgon
50
PulsedArgon Fast
40
Mode
Power
SoftCoag
40
PulseCut Slow
50
PulseCut Fast
40
ForcedCoag
40
ForcedArgon
35
CONTINUED ON NEXT PAGE
Effect
4
N/A
N/A
N/A
N/A
N/A
Effect
3
2
2
3
N/A