Olympus MU-1 Anleitung Handbuch - Seite 19

Blättern Sie online oder laden Sie pdf Anleitung Handbuch für Medizinische Ausrüstung Olympus MU-1 herunter. Olympus MU-1 36 Seiten. Maintenance unit
Auch für Olympus MU-1: Wartungshandbuch (28 seiten)

3.3
Preparation and Inspection of the Endoscope
and Disinfectant Container
Leakage Testing
Remove water from the Air/Water Channel (For endoscope
with Air/Water channel only)
MAINTENANCE UNIT MU-1
The Fiberscope Adapter (AC10-S) is not waterproof. Ensure
it is removed before immersing the OES Fiberscope in a
solution.
Refer to the instruction manuals of the Leakage Tester and endoscope for
directions on Leakage Testing.
1.
When using the OES Fiberscope, remove the Fiberscope Adapter from the
Light Guide Connector (if used).
2.
Remove the Air/Water Valve from the endoscope and replace it with the AW
Channel Cleaning Adapter (see Figure 3.4).
Figure 3.4
Chapter 3 Preparation and Inspection
15