Chapter 6: Safety Check

• Criteria
External leak currents should not exceed the values indicated below.
Inspection item

External leak current

• Inspection procedure
Measure the leak current at the external screw that is farthest from the inlet.
See the inspection procedures for leak current testing instrument for the measurement
methods.

2. Grounding Leak Current

• Criteria
Grounding leak currents should not exceed the values indicated below.
Inspection item
Grounding leak current
• Inspection procedure
Measure the leak current of the inlet grounding terminal.
See the inspection procedures for leak current testing instrument for the measurement
methods.
ISSUE6
1. External Leak Current
Normal condition
Normal condition
Criteria
Single fault
100 A
Criteria
500 A
6-1
500 A
Single fault
1000 A
Safety Check
MU-1