Warning Message
Change Fluid Bag
Deficit Reached
Overpressure
Deficit Exceeded
Critical Overpressure
Rapid Deficit
6
Hy s te r o flo w Qu i ck R ef e re n c e G u i d e
Begin the procedure
Warnings
Alert
Cause
3 Beeps
> 500 ml remaining
3 Beeps
Deficit exceeds the preset value
3 Beeps
Actual pressure 10 mmHg above nominal pressure
10 Beeps
In increments of 100 ml above the preset value
Actual pressure > 200 mmHg above nominal
10 Beeps
pressure
10 Beeps
Deficit change > 300 ml in 1 minute