6) In the procedure 1) to 3) in the drawing below, open the cover and take out the cantilever strip with
tweezers.
If you find the cantilever strip attached to the back side of the cover, stop opening the cover. Close
the case again and pat the cover with your finger to move the cantilever strip from the cover to the
bottom cover. Then try to open the case again.
Caution
Please handle our cantilevers carefully because they are fragile.
Caution
Avoid any contact with the cantilevers when you pull up the cantilever
strip from the case.
Caution
Cantilever chips should always be placed tip-side-up.
1) in the fig. : Press the hold position on the bottom cover to keep the stationary.
2) in the fig. : Using your thumb, carefully open the lid.
3) in the fig. : Pull the cantilever strip straight up with the tweezers.
Lip
1) Press the hold position on the bottom
cover to keep the case stationary.
The way to open the case and to put out the cantilever strip from the case
2) Using your thumb, carefully open the lid.
Tweezers
3) Pull the cantilever strip straight
up with the tweezers.
Warning)
Pay attention not to touch the
cantilevers to the case and the
other strips.
- 5e -
3
Preparation before disconnecting the cantilever chip
Caution
Do not use the plastic case for transportation as a work plate in the
process for chip disconnection.
1) Use a microscope slide glass as a work plate in the process for disconnecting the cantilever chip.
2) Lay the cantilever strip tip-side-up on the microscope slide glass.
3) Then, settle the cantilever strip on the microscope slide glass with a piece of Scotch drafting tape (below).
4
Disconnecting the cantilever chip
1) Disconnect the cantilever chip as follows.
Warning
Use protective eyewear when handling to avoid
damage to the eyes from breakage of the cantilever
chips.
Warning
Use protective gloves to avoid getting hurt by a shard of cantilever chip,
when breaking and cutting cantilever chips from cantilever strips.
- 6e -