LAMP HOUSING INSPECTION SHEET

{ Study the instruction manual for the lamp housing before inspection.
{ For safe use of the lamp housing, we recommend performing the following inspection periodically (every time you replace
the lamp and at least every 6 months).
{ The table below identifies the check items to be observed. Put (X) if not applicable or ( ) if applicable.
{ If there is any ( ) mark noted, immediately stop use of the product, and contact Olympus for detailed inspections or replace
the lamp housing.
{ If you detect an abnormality other than that listed below or with other Olympus product, also stop the use of the product and
contact Olympus for detailed inspections.
{Note that service, replacement and detailed inspections are charged after expiration of the warranty period.
1. More than 8 years have passed since original purchase or the total power ON
time has exceeded 20,000 hours.
2. Illumination flickers when you move the lamp cable or lamp housing.
3. Lamp cable is unusually hot to the touch.
4. Scorching or burning odor is produced during use.
5. Deformation, backlash, or looseness, etc. when you assemble the lamp housing.
(Impossibility of removing the top section of lamp housing when you attempt
to replace the lamp, etc.)
6. Discoloration, deformation or cracking of the lamp housing.
7. Melting, crack, deformation or solidification of the lamp cable or a wiring part.
8. Increased frequency of servicing compared to similar devices put into use at
the same time as the lamp housing.
* When the Check Result columns become insufficient, copy this sheet.
12
Check items
If you have any questions, please contact Olympus.
Check results (Date)
/
/
/
/