Guidance and manufacturer's declaration – electromagnetic emissions
The MODEL ABP801 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MODEL ABP801 should assure that it is used in such an
environment.
Emissions
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The MODEL ABP801 is intended for use in the electromagnetic environment specified below. The
customer or the user of the MODEL ABP801 should assure that it is used in such an environment.
Immunity test
IEC
level
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)IEC
61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines and
IEC 61000-4-4
patient coupled
lines
Surge
±1 kV line(s)
IEC 61000-4-5
and neutral
Voltage
dips,
<5 % UT
short
(>95 % dip in U ) T
interruptions
for 0,5 cycle
and voltage
40 % UT
variations on
(60 % dip in U ) T
power supply
for 5 cycles
input lines
70 % UT
IEC
(30 % dip in U ) T
61000-4-11
for 25 cycles
Complianc
Electromagnetic environment-- guidance
e
Group 1
The MODEL ABP801 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The MODEL ABP801 is suitable for use in all
establishments, including domestic
Class A
establishments and those directly connected to
the public low voltage power supply network that
Complies
supplies buildings used for domestic purposes.
60601
test
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines and
patient
lines
±1 kV line(s)
and neutral
<5 % UT
(>95 % dip in U ) T
for 0,5 cycle
40 % UT
(60 % dip in U ) T
for 5 cycles
70 % UT
(30 % dip in U ) T
for 25 cycles
Electromagnetic
environment--guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Main power quality should be that
of a typical commercial or hospital
coupled
environment.
Main power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If a dip or an
interruption of mains power occurs,
the current of the MODEL ABP801
may be dropped off from a normal
level. It may be necessary to use
an uninterruptible power supply or
a battery.
11