Guidance and manufacturer's declaration – electromagnetic emissions
The MODEL ABP802 is intended for use in the electromagnetic environment specified below. The
customer or the user of the MODEL ABP802 should assure that it is used in such an environment.
Emissions
RF emissions CISPR
11
RF emissions CISPR
11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The MODEL ABP802 is intended for use in the electromagnetic environment specified below. The customer
or the user of the MODEL ABP802 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
±6 kV contact
Discharge
±8 kV air
ESD)IEC
61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines and
IEC 61000-4-4
patient coupled lines
Surge IEC
±1 kV line(s) and
61000-4-5
neutral
Voltage dips,
<5 % UT
short
(>95 % dip in U ) T
interruptions
for 0,5 cycle
and voltage
40 % UT
variations on
(60 % dip in U ) T
power supply
for 5 cycles
input lines
70 % UT
IEC
(30 % dip in U ) T
61000-4-11
for 25 cycles
<5 % UT
Complianc
Electromagnetic environment-- guidance
e
Group 1
The MODEL ABP802 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class B
The MODEL ABP802 is suitable for use in all
Class A
establishments, including domestic establishments and
those directly connected to the public low voltage power
Complies
supply network that supplies buildings used for domestic
purposes
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines and
patient coupled lines
±1 kV line(s) and
neutral
<5 % UT
(>95 % dip in U ) T
for 0,5 cycle
40 % UT
(60 % dip in U ) T
for 5 cycles
70 % UT
(30 % dip in U ) T
for 25 cycles
<5 % UT
12
Electromagnetic environment
--guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If a dip or an interruption of
main power occurs, the current of the
MODEL ABP802 may be dropped off
from normal level, it may be necessary to
use uninterruptible power supply or a
battery.