5.4. Sterilize
5.4.1 Gravity Displacement Autoclave Sterilization
1. Wrap the Cover in sterile linens for distribution using a sequential envelope fold. Include
a sterile indicator or sterile indicator tape. Two 18 in by 18 in sterile linens is sufficient for
each Cover.
First fold lengthwise, dividing it in thirds, with the metal portion pointing upwards. Then
fold upwards from the bottom, again into thirds.
Keep the metal portion pointing upwards.
Folding cover along the same lines after each use will create concentrated areas of wear
at crease lines, reducing the lifetime of the Cover.
2. Wrap the chuck adapter and chuck key together in sterile linens for distribution using a
sequential envelope fold. Include a sterile indicator or sterile indicator tape. Two 10 in by
10 in sterile linens are sufficient for the chuck adapter and key.
Ensure chuck is in a fully-open position.
3. Place packages in autoclave and run the autoclave using these sterilization settings:
TEMPERATURE
EXPOSURE TIME
DRY TIME
Arbutus Medical
16
121ºC
30 minutes
60 minutes