Arc4Sports is only for use as directed.
Specifications:
Arc4Sports unit
Battery
Li-ion Rechargeable, 3.7V (fully traceable) - non removable
Pad connection
Via 2.1 x 5.5 mm (centre positive) plug on unit
Maximum Pad Voltage
+/- 9 V
Programme selection
Via push buttons
4 programmes
Pulse duration
Modes
Pulse to pulse spacing
Pulse polarity
Output channels
Single channel
0 - 9 V (+/-),
Max = 4V
Output voltage
Max = 0.2V
Output current
0 - 400 uA
Tolerances
All output parameters less than +/- 10%
Physical Dimensions:
45 mm (W) x 15 mm (D) x 105 mm (L)
Approximate weight:
Operating Temperature
0ºC to 40ºC
Storage Temperature
0ºC to 55ºC
Relative humidity
10 to 65%
IP rating
IP 3X
Expected life of unit
5 years under normal usage
DC Power Supply
Via 2.1 x 5.5 mm (centre positive) plug on unit
Connection
(pads removed whilst charging)
Mains supply
100-240VAC, 50-60Hz, 0.15A
Output supply
12VDC, 0.5A
Compliance
EN 60950
14
0.11 msec
0.86 msec
bipolar
controlled to maintain the specified current
for R = 10kohm (typical)
for R = 500 ohms
65 g
Warranty
Arc4Sports offers a 2-year guarantee on the Delivery Unit, commencing on the date of purchase.
In addition to your statutory rights, the manufacturer, Applied Microcurrent Technology Ltd agrees
that if any defect in materials or workmanship appears in this product within two years after the
original date of consumer purchase, it will be repaired or if necessary, replaced, free of charge.
This applies only if the product has been used for domestic purposes and has not been damaged
through misuse, accident or neglect and has not been modified or repaired by anyone other than
the manufacturer or its authorised agents.
If a defect appears, please make sure that the unit is being used in accordance with the
instructions. If so, return it with this warranty and the proof of purchase to Arc4Sports.
Exclusions: the Electrode Pads, Arm Strap and Plug Charger are not covered by this warranty –
only the Delivery Unit.
Electromagnetic Compatibility: Precautions
The equipment complies with the requirements of EN 60601-1-2, however, it should be noted,
hence it is recommended that close proximity to such devices (WiFi hubs, mobile and cordless
telephones, etc) be restricted during treatments.
WEEE Statement
The symbol on the product indicates that it should not be disposed of with your
normal household waste but should be taken to a designated collection point for
the recycling of waste electrical and electronic equipment.
Standard Symbols
Type BF Applied part
Attention
Complies with WEEE regulations
CE marking
Manufacturer
15