Siemens SITRANS WW310 Betriebsanleitung - Seite 33

Blättern Sie online oder laden Sie pdf Betriebsanleitung für Industrielle Ausrüstung Siemens SITRANS WW310 herunter. Siemens SITRANS WW310 48 Seiten. Weighfeeders

Siemens SITRANS WW310 Betriebsanleitung
8.7.3

Calibration

Step 1
Power up the integrator and using the appropriate integrator menu, verify that the new load
cell(s) are operating correctly.
Step 2
Zero calibrate the unit using the integrator manual for reference. Record zero calibration
procedure results.
Step 3
Span calibrate the unit using the integrator manual for reference. Record span calibration
procedure results.
Step 4
Re-check zero calibration. Run another zero calibration and record the results. Verify that the
zero repeated to the same value recorded in Step 2 (Page 31).
Step 5
Confirm calibration with material test as described in the integrator manual, if possible, before
placing the feeder back in operation.
SITRANS WW310
Operating Instructions, 03/2014, A5E33490433-AA
Start Up
8.7 Load cell replacement
31