2 - 4
2.2.7
Transverse movement drive
SIE
Limit switch
SIM
Travel range
•
Move the system over the entire travel range and check whether the limit switches are
activated:
- in the back tabletop position
(31.5 cm distance between the table rails and the rotation drive cover)
- in the front tabletop position (55.5 cm distance)
•
whether at a distance of 45.5 cm (in the "center" position)
- switch s61 responds and the tabletop stops
SIM
Ease of movement / play
•
Move the system over the entire travel range again and check for ease of movement and
play in the transverse direction.
2.2.8
Tabletop
SIM
Chains
•
Check that the chains are securely mounted.
SIM
Chain coupler
•
Check that the chain coupler on the tabletop is secure.
SIM
Safety distances
•
Raise the table to +88° and move it to maximum and minimum height.
•
Check the safety distance to the floor and to the ceiling (to the lights or to other ceiling-
mounted devices).
•
Move the tabletop in the horizontal position.
•
Insert the 87 cm long tabletop extension on the physician's side.
•
Move the table on the physician's side to the maximum end stop.
•
Check the safety distances to the wall or to other fixed installations.
•
Insert the tabletop extension on the head end.
•
Move the table headward to the maximum end stop.
•
Check the safety distances to the wall or to other fixed installations.
UROSKOP D
Inspection and maintenance
RLL5-310.101.02
Page 4 of 16
Rev. 02
01.97
TDSP 1
Siemens AG
Medical Engineering