Use the KV + or KV – switch to move to any program (PR 1 to PR7)
Use the DL_Serv switch on the D936 to to select the displayed program.
7.17.1 Program 1
Not Used
1. Short the ST link on the D915 and put the unit on .
the display will read as below and enter program mode .
2. This program is not used
7.17.2 Program 2
To check the total number of exposures taken .
1.The display on the D932 will be as below.
XXX XXX corresponds to the total number of exposures taken on the unit.
7.17.3 Program 3
To check the type of codes that occurred in the unit (maximum 20) .
1.Press the switch DL-serv on the D936 once.The display will be
AA diplays the serial number of the exposures that have occured, in chronological
order (max. 20) & XX displays the corresponding codes occured on the unit
7.17.4 Program 4
To erase the codes that occurred in the unit
1.Press the switch DL-serv on the D936 once.The display will be
Keep the mAS – key depressed for 4 seconds till the dislay beomes
7.17.5 Program 5
To store the Last Value of KV & mAS Display
Siemens Ltd. Med India
Pr
1
XXX
XXX
AA
XX
ErA
Err
0
Err