Attach the new collimator, fasten it with the four screws & follow the light field and
radiation field alignment procedure.
Connect the connecting cable.
Check coincidence of light and radiation field and make adjustment if required
8.13 Replacing the Collimator lamp
CAUTION
Screw down the bracket and the front panel.Check the function of the lamp.
Siemens Ltd. Med India
Loosen the screw of the clamp which has
marked for light indication on the collimator.
take out the defective lamp and replace it by a
new lamp.
Do not touch the glass envelope with your
bare fingers.