ACD CAL-101 Gebrauchsanweisung - Seite 5

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Tragbarer Generator ACD CAL-101 herunter. ACD CAL-101 15 Seiten. Calibration gas generator

ON/OFF Switch
The CAL-101 has no on/off switch. To use the instrument simply insert the
cell. To turn the unit off, simply remove the cell by pushing the cell eject
button located on the top of the instrument. When the unit is configured as
a 'bump' station, a modified generating cell is used. The 'bump' generating
cell is equipped with a switch to initiate a 'bump' cycle. In this configuration
the cell remains in the instrument and automatically turns off after a 30
second 'bump' cycle.
Delivery Hose
The instrument comes standard with a hose for delivering the gas to the
sensor calibration adapter. The air flow of the instrument is affected by the
hose length (and subsequent pressure drops through the hose) and must
be readjusted if alternative delivery hoses are used.
Page 3